MedicinePhysiologyRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyMagnetic Resonance ImagingMechanical VentilationPulmonary Function AssessmentPulmonary ImagingPulmonary Nuclear ImagingBackground NoiseBackground Noise EffectsLongitudinal RelaxationLongitudinal Relaxation TimeNuclear ResonancesPostprocessing TechniquesRelaxation Time MeasurementsSystematic Errors
A method for in vivo measurement of oxygen partial pressure (pO2) in porcine lungs during partial liquid ventilation (PLV) with perflubron (PFOB) was developed. A pulse sequence for high-resolution MRI of the distribution of PFOB in the lung after intratracheal administration was developed as well. Moreover, quantitative measurements of longitudinal relaxation time T(1) of (19)F resonances for assessment of regional pO2 are described. Due to the need to acquire data during a single expiratory breathhold, only low SNRs were achieved in vivo. Therefore, simulations were performed to investigate the influence of background noise on T(1) values calculated from data with low SNR. Based on these simulations, a postprocessing strategy was developed to correct for systematic errors by background noise prior to quantitative analysis. Results of a pilot study in pigs under conditions of PLV are presented.
24
<sup>1</sup>H NMR chemical shift selective (CHESS) imaging
Axel Haase, Jens Frahm, Wolfgang Hänicke et al. · Physics in Medicine and Biology · 1985 · 1K citations