Canadian Journal of Chemistry · 1967 · 11 citations · 0 references
Natural SciencesChemistryPhysicsAnalytical ChemistryAtomic PhysicsMolecular PhysicsDetector SystemsGas SensorsIonization ProcessesIonization TechniquesMass SpectrometryGas Detector PhysicsGas MixturesIonizationLock-in AmplificationMass Ratio DeterminationMolecular Beam ExperimentsPhase-sensitive DetectionSignal Formation
A method is described by which an ion gauge, used as a molecular-beam flux detector, can discriminate between signals produced by two gases, provided they have a mass ratio of at least 2 and are present in comparable amounts. The method employs a mechanically chopped beam with narrow band phase-sensitive detection. Separation of two signals is possible because of the different rates of effusion of the two gases out of the gauge. Because of this the phase of the a.c. signal extracted by the lock-in amplifier depends on the beam composition. A theoretical analysis of the phenomena is presented and is compared with experimental data. Limitations of the method are also considered.