We demonstrated that BMI does not affect the step output of commonly used activity monitors during walking. In addition, 67 m·min⁻¹ seems to be the minimum speed required for accurate step counting, at least for most waist-mounted activity monitors. Finally, the StepWatch, AG7164, and activPAL™ were the most accurate devices on the TM, but only the AG7164 yielded comparable step counts to the StepWatch in the free-living environment.