Development of a palm portable mass spectrometer

Mo Yang, Tae Young Kim, Hyun-Chul Hwang, Seok-Kyung Yi, Dohoon Kim

Journal of the American Society for Mass Spectrometry · 2008 · 129 citations · 12 references

Concepts

TL;DR

The palm‑portable mass spectrometer weighs 1.48 kg, measures 1.54 L, and operates on a 5 W battery using a miniaturized ion‑trap analyzer with four coaxial disks, 1500 Vpp/3.9 MHz RF, an ion‑getter pump, pulse‑mode gas introduction, and an embedded microcomputer for control. The device detected toluene at 6 ppm and dimethyl methylphosphonate at 22 ppm, indicating its potential as a personal mobile detector for chemical warfare agents.

Abstract

A palm portable mass spectrometer (PPMS) has been developed with a weight of 1.48 kg (3 lb) and a size of 1.54 L (8.2 x 7.7 x 24.5 cm(3)) that can be operated with an average battery power of 5 W. A miniaturized ion trap has been used as a mass analyzer that consists of four parallel disks with coaxial holes. A rf voltage of 1500 V(p-p) at 3.9 MHz has been used for scanning ion mass of up to m/z 300. An ion-getter pump serves for high vacuum of the PPMS. Sample gas was introduced in pulse mode. An embedded microcomputer has been developed for system control. Detection of organic gases diluted in the air has been demonstrated up to 6 ppm for toluene and 22 ppm for dimethyl methylphosphonate (DMMP). Performance results suggest usefulness of the PPMS as a personal mobile device for detection/identification of chemical warfare agents in the field.

References

12