Mammography of a phantom and breast tissue with synchrotron radiation and a linear-array silicon detector.

F. Arfelli, V. Bonvicini, Alberto Bravin, G. Cantatore, E. Castelli, L. Dalla Palma, Marco Di Michiel, Renata Longo, Alessandro Olivo, S. Pani,

Radiology · 1998 · 47 citations · 0 references

Concepts

Abstract

A linear-array, silicon pixel detector, capable of counting single photons, was applied to mammography by using a synchrotron radiation beam. Images were obtained of both a mammographic phantom and a breast-tissue sample. The phantom image was acquired with a mean glandular dose of 0.32 mGy. This detector combined with a synchrotron radiation beam allows acquisition of high-contrast, low-dose images of soft tissues.