3He Cryostat with Adsorption Pumping

C. F. Mate, R. F. Harris-Lowe, W. L. Davis, J. G. Daunt

Review of Scientific Instruments · 1965 · 32 citations · 2 references

Concepts

Abstract

A 3He cryostat is described in which temperatures down to 0.3°K have been obtained by pumping the 3He refrigerant bath with activated charcoal at a temperature of 1.3°K. The performance characteristics of the adsorption pump are described in some detail, and a discussion is given of the extent to which the speed, adsorptive capacity, and cooling requirements of such a pumping system can be predicted by calculation.

References

2