Journal of Biomedical Materials Research · 1989 · 58 citations · 15 references
Natural SciencesMaterials EngineeringMaterials ScienceMechanical Behavior Of MaterialsMechanical Properties Of Dental MaterialsStructural MaterialsBone CementsBone MechanicsFatigueFracture MechanicsStrength Of MaterialsBone Crack GrowthCrack GrowthFatigue AssessmentFatigue LifeFatigue Life AssessmentFatigue TestingFatigue Under Cyclic LoadingTemperature Effects On Fatigue
This study was conducted to investigate some fundamental fatigue testing variables as they apply to the response characteristics of acrylic bone cement. Cyclic loading under load control was conducted at frequencies of 1, 2, 5, 10, and 20 Hz in air at room temperature. At a tensile stress range of 0.3-20.0 MPa the fatigue life increased linearly with logarithmic frequency. The effect of conditioning and testing in saline at both room temperature and 37 degrees C at similar stress levels and a frequency of 10 Hz were also examined. When compared to dry testing at room temperature, testing in saline at 37 degrees C resulted in a reduction in fatigue life while testing in saline at room temperature produced an increase in fatigue life. Of a number of statistical distributions considered, the Weibull was found to be the most appropriate in documenting the findings of this investigation. A companion fractographic investigation of the failure surfaces demonstrated distinct regions of crack growth and fast fracture.
15
Robert D. Beckenbaugh, D M Ilstrup · Journal of Bone and Joint Surgery · 1978 · 362 citations
Carbon fiber‐reinforced bone cement in orthopedic surgery
R. M. Pilliar, R.A. Blackwell, Ian Macnab et al. · Journal of Biomedical Materials Research · 1976 · 125 citations