Flexible screen printed biosensor with high-Q microwave resonator for rapid and sensitive detection of glucose

Kishor Kumar Adhikari, Zorigt Chuluunbaatar, Hyejin Park, Younsu Jung, Gyoujin Cho, Yong Hwa Jo, Sung Soo Kim, Nam‐Young Kim

2014 · 15 citations · 6 references

Concepts

Abstract

This paper presents a rapid and sensitive mediator-free glucose biosensor based on microwave resonator implemented using circularly folded T-shaped uniform impedance resonators, screen printed on flexible polyethylene substrate. As a result of high-Q factor of 160, the proposed glucose biosensor features high sensitivity and ultralow detection limit of 71 MHz/mgmL <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-1</sup> and 0.0167 μM, respectively, at a central frequency of 11.8 GHz, within linear detection range of 1 to 5 mg/mL. Additionally, the clear dependence of loaded quality factor (Q <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">L</sub> ), return loss (S <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">11</sub> ), propagation constant (γ), and impedance (Z) on glucose level enables the effective multidimensional detection of glucose sensor.

References

6