2013 · 14 citations · 7 references
EngineeringElectrical EngineeringAnalog ElectronicsBiomedical ImagingCircuit TheoryImpedance MatchingInstrumentation AmplifiersCole-cole ModelElectrical ImpedanceElectrical Impedance TomographyImpedance MeasurementOutput ImpedanceOutput Impedance AnalysisSource Impedance Modeling
An accurate current source is one of the keys in the hardware of Electrical impedance Tomography systems. Limitations appear mainly at higher frequencies and for non-simple resistive loads. In this paper, we simulate an improved Howland current source with a Cole-Cole load. Simulations comparing two different op-amps (THS4021 and OPA843) were performed at 1 kHz to 1 MHz. Results show that the THS4021 performed better than the OPA843. The current source with THS4021 reaches an output impedance of 20 MΩ at 1 kHz and above 320 kΩ at 1 MHz, it provides a constant and stable output current up to 4 mA, in the complete range of frequencies, and for Cole-Cole (resistive and capacitive) load.
7
Electrical Impedance Tomography
John G. Webster, Paul D. Hughes · Journal of Clinical Engineering · 1991 · 218 citations
Current source design for electrical impedance tomography
Alexander S. Ross, G.J. Saulnier, J.C. Newell et al. · Physiological Measurement · 2003 · 160 citations
Engineering, Biomedical Engineering, Electrical Engineering +10
Precision constant current source for electrical impedance tomography
Jeong‐Whan Lee, Tong In Oh, Sang Min Paek et al. · 2004 · 28 citations
Engineering, Biomedical Engineering, Electrical Engineering +16