Publication | Closed Access
Comparison of Intraocular Pressure Measurements and Assessment of Intraobserver and Interobserver Reproducibility With the Portable ICare Rebound Tonometer and Goldmann Applanation Tonometer in Glaucoma Patients
65
Citations
20
References
2013
Year
RBT may have a potential role and use in routine evaluation and management of glaucoma patients given its high correlation with GAT and high reproducibility in IOP measurements. As with GAT, measurements obtained with RBT are affected by CCT, reinforcing the role of pachymetry in glaucoma management.