Analytical Chemistry · 1993 · 232 citations · 19 references
EngineeringBiochemistryBioengineeringAnalytical ChemistryBiosensorsElectroanalytical MethodsBioanalytical ChemistryBiosensors For Immune DetectionMicroscale BiosensorsPiezoelectric BiosensorsProtein AdsorptionBovine Serum AlbuminPiezoelectric BiosensingPiezoelectric ImmunosensorsPiezoelectric Protein BiosensorsQuartz Crystal MicrobalanceQuartz Crystal Microbalance BiosensorsQuartz Crystal Microbalance ImmunosensorsQuartz Crystal Microbalance Sensors
We have developed a piezoelectric immunosensor for the detection of microalbumin. Human serum albumin (HSA) in the range 0.1-100 micrograms mL-1 could be detected using a flow cell; the immunosensor is sensitive enough to monitor levels of albuminuria. The immunosensor did not respond to bovine serum albumin, only to HSA, implying that the specificity for HSA was high. We investigated the relationship between the frequency change (delta F) and adsorption per unit area of piezoelectrically active quartz crystal (delta M). delta M was estimated with radioisotope-labeled anti-HSA or HSA. When anti-HSA was adsorbed onto the surface of the crystal or HSA was bound to anti-HSA supported by the crystal, values of magnitude of delta F/delta M were larger than the value predicted from theory (Sauerbrey's equation). Furthermore, magnitude of delta F/delta M for HSA was larger than that for anti-HSA.
19