Multi–Detector Row CT Colonography: Effect of Collimation, Pitch, and Orientation on Polyp Detection in a Human Colectomy Specimen

Stuart A. Taylor, Steve Halligan, Clive I. Bartram, P R Morgan, Ian Talbot, Nicola Fry, Brian P. Saunders, Kirosh Khosraviani, Wendy Atkin

Radiology · 2003 · 67 citations · 19 references

Abstract

Detection of small polyps (<5 mm) with multi-detector row CT is highly dependent on collimation, pitch, and, to a lesser extent, tube current. Collimation of 1.25 mm, combined with pitch of 6 and tube current of 50 mA, provides for reliable detection of polyps 5 mm or larger while limiting the effective radiation dose. Polyps smaller than 5 mm, however, may be poorly depicted with use of these settings in the transverse colon.