IEEE Transactions on Instrumentation and Measurement · 1972 · 27 citations · 1 references
EngineeringElectrical EngineeringAnalog ElectronicsInstrumentationPower ElectronicsElectrical CharacterizationElectrical MeasurementElectrical MetrologyEnergy MeteringPower QualityPower TestingElectrical InstrumentsElectrical Measurement SystemsElectrical Power QualityElectronic InstrumentationPower FactorPower MeasurementPrecision MeasurementVoltage Measurement
The accuracy and frequency response limitations in the present square-law responding laboratory wattmeters have promoted the search for alternate methods of ac power measurement using electronic instrumentation. The three-voltmeter method is based on an old principle of operation implemented by new analog circuitry. Results of the tests reported here show a precision of power calculation with 0.01 percent error at frequencies below 5 kHz. A unique test procedure is described for the comparison of the three-voltmeter device and a time-division multiplier wattmeter operating at zero power factor.
1