Evaluation of a new tracheal tube with a movable bronchus blocker

R. G. MacGillivray

Anaesthesia · 1988 · 44 citations · 3 references

DOIFull text

Open access

Concepts

Abstract

A new tracheal tube which incorporates a movable bronchus blocker (the Univent tube), was evaluated in eight patients who underwent thoracic surgery. Difficulties encountered with accurate placement suggest that routine use should be made of a fibreoptic bronchoscope to confirm the position of the blocker.

References