Investigative Radiology · 2009 · 76 citations · 45 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingBiomedical Signal ProcessingDiagnostic ImagingCardiac ImagingCardiovascular ImagingElectrocardiographyMagnetic Resonance ImagingMedical Imaging InstrumentationMedical InstrumentationCardiac Cine ImagingCine Magnetic Resonance ImagingElectromagnetic InterferenceElectromagnetic Interference MitigationHigh-field Magnetic Resonance ImagingMagnetic Resonance Imaging InstrumentationMedical Signal ConditioningPhysiological Signal Processing
OBJECTIVES: To circumvent the challenges of conventional electrocardiographic (ECG)-gating by examining the efficacy of an MR stethoscope, which offers (i) no risk of high voltage induction or patient burns, (ii) immunity to electromagnetic interference, (iii) suitability for all magnetic field strengths, and (iv) patient comfort together with ease of use for the pursuit of reliable and safe (ultra)high field cardiac gated magnetic resonance imaging (MRI). MATERIALS AND METHODS: The acoustic gating device consists of 3 main components: an acoustic sensor, a signal processing unit, and a coupler unit to the MRI system. Signal conditioning and conversion are conducted outside the 0.5 mT line using dedicated electronic circuits. The final waveform is delivered to the internal physiological signal controller circuitry of a clinical MR scanner. Cardiovascular MRI was performed of normal volunteers (n = 17) on 1.5 T, 3.0 T and 7.0 T whole body MR systems. Black blood imaging, 2D CINE imaging, 3D phase contrast MR angiography, and myocardial T2* mapping were carried out. RESULTS: The MR-stethoscope provided cardiograms at 1.5 T, 3.0 T and 7.0 T free of interference from electromagnetic fields and magneto-hydrodynamic effects. In comparison, ECG waveforms were susceptible to T-wave elevation and other distortions, which were more pronounced at higher fields. Acoustically gated black blood imaging at 1.5 T and 3.0 T provided image quality comparable with or even superior to that obtained from the ECG-gated approach. In the case of correct R-wave recognition, ECG-gated 2D CINE SSFP imaging was found to be immune to cardiac motion effects -even at 3.0 T. However, ECG-gated 2D SSFP CINE imaging was prone to cardiac motion artifacts if R-wave mis-registration occurred because of T-wave elevation. Acoustically gated 3D PCMRA at 1.5 T, 3.0 T and 7.0 T resulted in images free of blood pulsation artifacts because the acoustic gating approach provided cardiac signal traces free of interference with electromagnetic fields or magneto-hydrodynamic effects even at 7.0 Tesla. Severe ECG-trace distortions and T-wave elevations occurred at 3.0 T and 7.0 T. Acoustically cardiac gated T2* mapping at 3.0 T yielded a T2* value of 22.3 +/- 4.8 ms for the inferoseptal myocardium. CONCLUSIONS: The proposed MR-stethoscope presents a promising alternative to currently available techniques for cardiac gating of (ultra)high field MRI. Its intrinsic insensitivity to interference from electromagnetic fields renders it suitable for clinical imaging because of its excellent trigger reliability, even at 7.0 Tesla.
45
Andrew C. Larson, Richard D. White, Gerhard Laub et al. · Magnetic Resonance in Medicine · 2003 · 412 citations · Full text
Medicine, Bioengineering, Radiology +17
Preliminary report on in vivo coronary MRA at 3 Tesla in humans
Matthias Stuber, René M. Botnar, Stefan Fischer et al. · Magnetic Resonance in Medicine · 2002 · 224 citations · Full text
Medicine, Cardiology, Radiology +17
Phonocardiogram signal analysis: a review.
Rangaraj M. Rangayyan, Richard J. Lehner · PubMed · 1987 · 199 citations
Medicine, Cardiology, Physiology +13