Artificial Organs · 1999 · 34 citations · 3 references
EngineeringNeuroscienceRehabilitation MedicineMotor ControlMovement AnalysisNeural EngineeringFoot RehabilitationFunctional Electrical StimulationNeural StimulationNeurostimulationAdaptive Logic ModelsControl Signal GenerationDrop Foot CorrectionLow-pass FilteringLower Limb Functional Electrical StimulationSensory Feedback InterfacesSensory Feedback LoopsStance Phase Control
We report on our advances in sensory feedback data processing and control system design for functional electrical stimulation (FES) assisted correction of foot drop. We have applied 2 methods of signal purification on the bin integrated electroneurogram (i.e., optimized low pass filtering and wavelet denoising) before training adaptive logic networks (ALN). ALN generated stimulation control pulses, which correspond to the swing phase of the impaired leg when dorsal flexion of the foot is necessary to provide safe ground clearance. However, the obtained control signal contained sporadic stimulation spikes in the stance phase, which can collapse the subject, and infrequent broken stimulation pulses in the swing phase, which can result in unpredictable consequences. In this study, we have introduced adaptive restriction rules (ARR), which are initially used as previously reported and then dynamically adapted during the use of the system. Our results suggest that ARR provide a safer and more reliable stimulation pattern than fixed restriction rules.
3