IEEE Transactions on Magnetics · 2012 · 25 citations · 9 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingDiagnostic ImagingElectromagneticsImagingMagnetic CharacterizationMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingSignal Detection TheoryExcitationMagnetic Field HarmonicsMagnetic ImagingMagnetic Particle ImagingMagnetization CurvesSignal-to-noise RatioSignal-to-noise Ratio AnalysisSingular Value Decomposition
We developed a magnetic nanoparticle (MNP) imaging system that uses the harmonic signals of MNPs on the basis of their nonlinear magnetization curves. By measuring the harmonic signals, the interference of the excitation field can be significantly decreased, so the detection sensitivity can be improved. First, we obtained a contour map of the signal field from the MNPs, from which we estimated their positions. We could detect 100 μg of MNPs located at <i xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">z</i> =30 mm below the pickup coil with a signal-to-noise ratio of 4. Next, we analyzed the contour map using a mathematical technique called singular value decomposition (SVD) in order to improve the spatial resolution of the estimated position. Using the SVD method, we converted the field map to the MNP distribution. We could clearly distinguish two MNP samples separated by <i xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">x</i> =20 mm and located at Δ <i xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">z</i> =20 mm. This simple detection system will be useful for medical applications such as sentinel lymph node detection.
9
Single-sided device for magnetic particle imaging
Timo F. Sattel, Tobias Knopp, Sven Biederer et al. · Journal of Physics D Applied Physics · 2008 · 153 citations