Concepedia

Publication | Closed Access

A Comparison of Flow Rates and Warming Capabilities of the Level 1 and Rapid Infusion System with Various-Size Intravenous Catheters

63

Citations

13

References

2003

Year

Abstract

The rapid infusion system is superior to the Level 1 for warming and flow of crystalloid for IV catheters >18 gauge in vitro. The rapid infusion system provides no advantage with catheters typically used in small children (<or=18 gauge). Safety and cost are additional factors to be considered when choosing one system over the other.

References

YearCitations