Dosimetric properties of photon beams from a flattening filter free clinical accelerator
213
Citations
14
References
2006
Year
Accelerator PhysicRadiation Therapy PlanningInstrumentationDepth Dose DependenciesRadiation OncologyNuclear MedicineAccelerator TechnologyRadiologyHealth SciencesRadiation TherapyMedical ImagingCosmic RayProton TherapyPhoton BeamsDosimetryWater PhantomDosimetric PropertiesRadiation DoseVarian Clinac 21ExMedicineParticle AcceleratorBeam Transport System
Basic dosimetric properties of 6 MV and 18 MV photon beams from a Varian Clinac 21EX accelerator operating without the flattening filter have been measured. These include dose rate data, depth dose dependencies and lateral profiles in a water phantom, total scatter factors and transmission factors of a multileaf collimator. The data are reviewed and compared with measurements for the flattened beams. The unflattened beams have the following: a higher dose rate by factors of 2.3 (6 MV) and 5.5 (18 MV) on the central axis; lower out-of-field dose due to reduced head scatter and softer spectra; less variation of the total scatter factor with field size; and less variation of the shape of lateral dose profiles with depth. The findings suggest that with a flattening filter free accelerator better radiation treatments can be developed, with shorter delivery times and lower doses to normal tissues and organs.