Publication | Open Access
Three-dimensional flow contrast imaging of deep tissue using noncontact diffuse correlation tomography
57
Citations
13
References
2014
Year
This study extended our recently developed noncontact diffuse correlation spectroscopy flowmetry system into noncontact diffuse correlation tomography (ncDCT) for three-dimensional (3-D) flow imaging of deep tissue. A linear array of 15 photodetectors and two laser sources connected to a mobile lens-focusing system enabled automatic and noncontact scanning of flow in a region of interest. These boundary measurements were combined with a finite element framework for DCT image reconstruction implemented into an existing software package. This technique was tested in computer simulations and using a tissue-like phantom with anomaly flow contrast design. The cylindrical tube-shaped anomaly was clearly reconstructed in both simulation and phantom. Recovered and assigned flow contrast changes in anomaly were found to be highly correlated: regression slope = 1.00, R<sup>2</sup> = 1.00, and p < 10<sup>-5</sup> in simulation and regression slope ≥ 0.97, R<sup>2</sup> ≥ 0.96, and p < 10<sup>-3</sup> in phantom. These results exhibit promise of our ncDCT technique for 3-D imaging of deep tissue blood flow heterogeneities.
| Year | Citations | |
|---|---|---|
Page 1