A gamma‐ray detector for capillary zone electrophoresis and its use in the analysis of some radiopharmaceuticals
28
Citations
4
References
1990
Year
EngineeringGamma‐ray DetectorCapillary Zone ElectrophoresisAnalytical InstrumentationAnalytical ChemistryClinical ChemistryIsotachophoresisNuclear MedicineBiophysicsChromatographyRadiologyGamma-ray DetectorCapillary ElectrophoresisRadiation DetectionChromatographic AnalysisMass SpectrometryRadioanalytical ChemistryLinear ResponseMedicineDrug Analysis
The construction and evaluation of a gamma-ray detector for capillary zone electrophoresis is described. The detector was shown to have a linear response from the limit of detection 10 to 500 Bqcm-3 corresponding to 5.1 x 10(-17) to 2.55 x 10(-15) gcm-3 43Tc99m. The application of the detector for the analysis of some radiopharmaceuticals is presented.