Analytical Communications · 1997 · 13 citations · 7 references
PharmacyDermatologyAnalytical ToxicologyChromatographic AnalysisMass Spectrometric AnalysisBioanalytical ChemistryBioanalytical Method ValidationChromatography CouplingInstrumental AnalysisMass SpectrometryVitamin AnalysisElectrospray IonizationHigh-performance Liquid Chromatography-mass SpectrometryIon Trap Mass SpectrometryLiquid Chromatography-mass SpectrometryLiquid Chromatography-mass Spectrometry Method DevelopmentMethod Detection LimitReversed-phase ColumnsSelected Ion Monitoring
A method for the determination of the vitamin biotin in extracts from different layers of human skin is presented, which has a detection limit of 800 pg ml–1 without further sample preparation. The quantification was performed using an HPLC procedure on a reversed phase column, which takes only 4 min, coupled to a single quadrupole mass spectrometer via an electrospray ionization interface. The detection was performed in the positive selected ion monitoring mode. Additional MS–MS and MSn experiments with biotin using an ion trap mass spectrometer are described.
7
Jörg Schiewe, Steffen Göbel, Maria Schwarz et al. · Journal of Pharmaceutical and Biomedical Analysis · 1996 · 23 citations
Pharmacy, Pharmaceutical Analysis, Pharmaceutical Biotechnology +14