Acoustic power calibrations of cylindrical intracavitary ultrasound hyperthermia applicators

Kullervo Hynynen

Medical Physics · 1993 · 77 citations · 0 references

Concepts

TL;DR

Preliminary clinical results indicate that some tumors can be heated effectively using cylindrical ultrasound sources placed in body cavities. This paper presents a simple method to measure acoustic power from cylindrical intracavitary transducers. The method converts the radially propagating acoustic field into a single‑direction beam using a brass reflector and measures the resulting radiation force on an absorbing target. The study found significant variation in power output among identical transducer elements, underscoring the need for individual element measurement, and demonstrated that the radiation force technique is simple, sensitive, and suitable for routine clinical quality assurance.

Abstract

Preliminary clinical results indicate that some tumors can be heated well utilizing cylindrical ultrasound sources placed in body cavities. In this paper a simple method for measuring the acoustic power from cylindrical intracavitary transducers will be described. The radially propagating acoustic field was converted to a beam with a single propagation direction by a brass reflector, and the radiation force generated by this beam on an absorbing target was measured. The power output of several clinical intracavitary arrays varied significantly between identically shaped transducer elements. The results show that it is important to measure the acoustic power output from each element prior to its clinical use. The radiation force technique is simple and sensitive and can be easily adapted to be used as a routine clinical quality assurance method.