Electrophoresis microchip fabricated by a direct‐printing process with end‐channel amperometric detection

Wendell K. T. Coltro, José Alberto Fracassi da Silva, Heron D. Torres da Silva, Eduardo M. Richter, Rogério Furlan, Lúcio Angnes, Claudimir Lúcio do Lago, Luiz Henrique Mazo, Emanuel Carrilho

Electrophoresis · 2004 · 65 citations · 24 references

Concepts

Abstract

We describe the development of an electrophoresis microchip fabricated by a direct-printing process, based on lamination of printed polyester films with end-channel amperometric detection. The channel structures are defined by polyester (base and cover) and by a toner layer (walls). The polyester-toner devices presented an electroosmotic flow (EOF) magnitude of approximately 10(-5) cm2 V(-1) s(-1), which is generated by a polymeric mixture of the toner and polyester composition. The microelectrodes used for detection were produced combining this laser-printer technology to compact discs. The performance of this device was evaluated by amperometric detection of iodide and ascorbate. The detection limits found were 500 nmol.L(-1) (135 amol) and 1.8 micromol.L(-1) (486 amol) for iodide and ascorbate, respectively.

References

24