The Effect of Peripherally Inserted Central Catheter (PICC) Valve Technology on Catheter Occlusion Rates - The ‘ELeCTRiC’ Study

Andrew J. Johnston, C. T. Streater, Remy Noorani, Joanne L. Crofts, Aldwin B. Del Mundo, Richard Parker

The Journal of Vascular Access · 2012 · 35 citations · 11 references

Concepts

Abstract

PURPOSE: Peripherally Inserted Central Catheters (PICCs) are increasingly being used to provide short to medium-term central venous access. The current study was designed to test the hypothesis that PICC valve technology does not influence PICC occlusion rates. METHODS: Intensive care unit (ICU) patients who required a PICC were randomized to one of three types of dual lumen PICC (open ended non-valved, Groshong valve, PASV valve). PICC occlusions were recorded and managed with a protocol that used urokinase. RESULTS: A total of 102 patients were recruited to the study. The overall risk of occlusion per catheter was 35% (95% CI 26% to 44%). The overall rate of occlusion was 76 occlusions per 1000 catheter days (95% CI 61 to 95). Presence or type of valve did not significantly influence this rate (open-ended non-valved PICC 38% of catheters, 79 occlusions per 1000 catheter days; Groshong 38% of catheters, 60 occlusions per 1000 catheter days; PASV 27% of catheters, 99 occlusions per 1000 catheter days). The dose of urokinase required to treat PICC occlusions did not significantly differ between PICC types. CONCLUSIONS: Valved PICCs do not appear to influence PICC occlusion rates.

References

11