Radiology · 1989 · 34 citations · 0 references
MedicineObstetricsRadiologyDiagnostic ImagingDiagnostic RadiologyRadiation DosimetryComputed TomographyComputed Tomography PhysicsFetal ImagingPhantom DosimetryPrenatal ImagingRadiation Dose MeasurementComputed Tomography Dose MeasurementComputed Tomography Phantom DosimetryFetal Dose LimitsFetal PelvimetryMedical Imaging Radiation DosePhantom-based Dose MeasurementRadiation Dose EstimationRadiographic Pelvimetry
Fetal and maternal dose estimates for computed tomographic pelvimetry have been obtained from phantom measurements. Use of routine abdomen imaging techniques may result in localized fetal doses in excess of 13 mGy (1.3 rad). With the use of a low-exposure (40-mAs) technique, it is possible to obtain images of acceptable quality for the necessary measurements. The resulting dose to the fetus is approximately 2.3 mGy (0.23 rad).