Publication | Open Access
Analysis of load sharing on uncovertebral and facet joints at the C5–6 level with implantation of the Bryan, Prestige LP, or ProDisc-C cervical disc prosthesis: an in vivo image-based finite element study
68
Citations
19
References
2010
Year
In unconstrained/semiconstrained prostheses with different core rigidity, the shared loads at the joints differ, and greater flexibility may result in greater joint loads. With respect to the 3 artificial discs studied, load sharing of the Bryan disc was highest and was closest to normal load sharing with the facet and uncovertebral joints. The Prestige LP and ProDisc-C carried more load through their rigid core, resulting in decreased load transmission to the facet and uncovertebral joints.