A multicenter measurement of magnetization transfer ratio in normal white matter

I. Berry, Gareth J. Barker, Frederik Barkhof, A. Campi, Vincent Dousset, Jean‐Michel Franconi, Achim Gass, Wolfgang Schreiber, David H. Miller, Paul S. Tofts

Journal of Magnetic Resonance Imaging · 1999 · 120 citations · 28 references

DOIFull text

Open access

Concepts

Abstract

To assess the importance of intercenter variations when measuring magnetization transfer ratio (MTR) in the brain, six European centers measured MTR in normal white matter. MTR ranged from 9 to 51 percent units (25 sequences). The effective flip angle of the saturating pulse divided by the pulse repetition time (ENRsat degrees/msec) was a good predictor of MTR (MTR = 3.25 ENRsat).

References