Diabetes Technology & Therapeutics · 2004 · 152 citations · 25 references
MedicineBiological EngineeringEndocrinologyDiabetologyMetabolic RegulationPhysiological Systems ModelingContinuous Glucose MonitoringDiabetes ControlDiabetes ManagementDiabetes MellitusGlycemic ControlGlycemic MonitoringArtificial Pancreas SystemsContinuous Subcutaneous Insulin InfusionControl Loop DesignFeasibility Study DesignGlucose ConcentrationInsulin PumpsIntravenous Insulin InfusionPredictive Glucose Modeling
The objective of the project Advanced Insulin Infusion using a Control Loop (ADICOL) was to develop a treatment system that continuously measures and controls the glucose concentration in subjects with type 1 diabetes. The modular concept of the ADICOL's extracorporeal artificial pancreas consisted of a minimally invasive subcutaneous glucose system, a handheld PocketPC computer, and an insulin pump (D-Tron, Disetronic, Burgdorf, Switzerland) delivering subcutaneously insulin lispro. The present paper describes a subset of ADICOL activities focusing on the development of a glucose controller for semi-closed-loop control, an in silico testing environment, clinical testing, and system integration. An incremental approach was adopted to evaluate experimentally a model predictive glucose controller. A feasibility study was followed by efficacy studies of increasing complexity. The ADICOL project demonstrated feasibility of a semi-closed-loop glucose control during fasting and fed conditions with a wearable, modular extracorporeal artificial pancreas.
25
[Lys(B28), Pro(B29)]-Human Insulin: A Rapidly Absorbed Analogue of Human Insulin
Daniel C. Howey, Ronald R. Bowsher, Rocco L. Brunelle et al. · Diabetes · 1994 · 451 citations
An Artificial Endocrine Pancreas
A. M. Albisser, B. S. Leibel, T G Ewart et al. · Diabetes · 1974 · 378 citations
Medicine, Endocrinology, Physiology +17