Medical Physics · 2000 · 47 citations · 20 references
EngineeringRadiologyRadiation DetectionRadiation DosimetryRadiation MeasurementRadiation Dose MeasurementRadiographic ImagingX-ray DetectionX-ray DetectorsX-ray OpticsDetector Spatial ResolutionGafchromic Film DosimetryHigh-resolution DosimetryLateral Dose ProfilesMicrobeam DosimetrySpatial Resolution MeasurementX-ray CollimatorsX-ray Dosimetry
Measurement of the lateral profile of the dose distribution across a narrow x-ray microbeam requires a dosimeter with a micron resolution. We investigated the use of a MOSFET dosimeter in an "edge-on" orientation with the gate insulating oxide layer parallel to the direction of the beam. We compared results using this technique to Gafchromic film measurements of a 200 micrometer wide planar x-ray microbeam. The microbeam was obtained by using a vernier micrometer-driven miniature collimator attached to a Therapax DXT300 x-ray machine operated at 100 kVp. The "edge-on" application allows utilization of the ultra thin sensitive volume of the MOSFET detector. Spatial resolution of both the MOSFET and Gafchromic film dosimeters appeared to be of about 1 micrometer. The MOSFET dosimeter appeared to provide more uniform dose profiles with the advantage of on-line measurements.