Publication | Closed Access
Optimal Power Settings for Holmium:YAG Lithotripsy
176
Citations
22
References
2012
Year
Holmium:YAG lithotripsy varies as pulse energy settings vary. At low pulse energy (0.2 J) less fragmentation and retropulsion occur and small fragments are produced. At high pulse energy (2.0 J) more fragmentation and retropulsion occur with larger fragments. Anti-retropulsion devices produce more efficient lithotripsy, particularly at high pulse energy. Optimal lithotripsy laser dosimetry depends on the desired outcome.