MedicineOrthopaedic SurgeryOrthopedic SurgeryReconstructive SurgerySpinal BiomechanicsSpine SurgeryFixation BiomechanicsFracture FixationFracture MechanicsFracture RepairInternal FixationVertebral FracturesBiomechanics Of Fixation ConstructsBiomechanics Of Fracture RepairFracture Fixation BiomechanicsSacroiliac FixationScrew FixationStatic StrengthStatic Strength AnalysisStatic Strength Testing
The relative static strengths of seven fixation configurations for the repair of sacroiliac fracture-separations were determined. The fixation techniques were two 3.5 mm screws, one 4.5 mm screw, one 4.5 mm screw and a pin, two 4.5 mm screws, one 6.5 mm screw, one 6.5 mm screw and a pin, and two 6.5 mm screws. Each configuration was subjected to a torsional, a bending, and a shear disruptive force while the maximum load tolerated before failure was measured. Two screws were stronger than a single screw of similar size, two small screws were stronger than a single larger screw, and a reduction pin added no significant strength to a single screw repair. It was concluded that use of the largest screws possible will maximize the strength of a sacroiliac fracture-separation repair.