Review of Scientific Instruments · 1967 · 26 citations · 1 references
EngineeringElectrical EngineeringInstrumentationRadiation DetectionRadiation MeasurementRadiation InstrumentationTemperature MeasurementTemperature MetrologyTemperature SensorsEmittanceExperimental Error ControlMetal Surface CharacteristicsNew Instrument DesignPolarimetric Error AnalysisPolarimetric Instrument DesignPolarimetryPolarimetry Instruments
A new automatic instrument for the measurement of surface temperatures is described. The instrument examines for polarization the sum of (1) radiation emitted by the test surface and (2) radiation emitted by a blackbody and reflected from the test surface. In measurements up to 450°C on metallic surfaces with emittances of 0.05 to 0.47, the determination of equality of sample and reference temperatures was accurate within about ±2%, with no emittance corrections. Further tests are required to apportion this error among sample surface characteristics, instrument characteristics, and experimental error. Away from the null, the error signal depends on the difference of the emittances εp and εs.
1
A. G. Worthing · Journal of the Optical Society of America · 1926 · 58 citations
Natural Sciences, Materials Science, Electronic Materials +11