Monitoring In Vivo Implant Loads With a Telemeterized Internal Spinal Fixation Device
133
Citations
17
References
2000
Year
The results show that fixators may be highly loaded even after fusion has occurred. A flexion bending moment acts on the implant even with the body in a relaxed lying position. This means that already shortly after the anterior procedure, the shape of the spine is not neutral and unloaded, but slightly deformed, which loads the fixators. Pedicle screw breakage more than half a year after insertion does not prove that anterior interbody fusion has not occurred.