EngineeringBiomedical EngineeringMicroelectromechanical SystemsRadiation DetectionRadiation MeasurementEnergy ResolutionRadiographic ImagingX-ray DetectionX-ray DetectorsX-ray OpticsX-ray SpectroscopyCurrent-voltage RelationsDetector Energy ResolutionImaging ResolutionMicrocalorimeter DetectorsRadiation Energy ResolutionSmall-signal ModelingX-ray Detector ArraysX-ray Imaging Performance
Frequency domain multiplexing (FDM) is an attractive option for the readout of imaging arrays of microcalorimeters. Implementation of FDM requires ac biasing of the individual microcalorimeters. In this letter we present a small signal model for the behavior of a microcalorimeter under ac bias. Moreover, we have measured the behavior of the same microcalorimeter under ac (at 46 kHz) and dc bias. These experiments show that the performance of the device is very similar in terms of energy resolution, pulse shapes, and current–voltage characteristics. The measured energy resolution at 5.89 keV is 6.9 eV for ac bias and 5.5 eV for dc bias. The effective time constant in both cases is 100 μs.
5
Superconducting multiplexer for arrays of transition edge sensors
J. A. Chervenak, K. D. Irwin, Erich N. Grossman et al. · Applied Physics Letters · 1999 · 283 citations