Magnetic Resonance in Medicine · 2007 · 224 citations · 9 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyCardiovascular ImagingMagnetic Resonance ImagingMagnetic Resonance Pulse SequencesMedical ImagingVascular AnatomyVascular ImagingArterial Spin LabelingCerebral Vascular TerritoriesMagnetic Resonance AngiographyMagnetic Resonance Flow ImagingMagnetic Resonance Flow Quantification
A new signal-to-noise ratio (SNR) efficient method is introduced for the mapping of vascular territories based on pseudocontinuous arterial spin labeling (ASL). A pseudocontinuous tagging pulse train is modified using additional transverse gradient pulses and phase cycling to place some arteries in a tag condition, while others passing through the same tagging plane are in a control condition. This is combined with a Hadamard or similar encoding scheme such that all vessels of interest are fully inverted or relaxed for nearly all of the encoding cycles, providing optimal SNR. The relative tagging efficiency for each vessel is measured directly from the ASL data and is used in the decoding process to improve the separation of vascular territories. High SNR maps of left carotid, right carotid, and basilar territories are generated in 6 min of scan time.
9
R. E. A. C. Paley · Journal of Mathematics and Physics · 1933 · 607 citations
Flow Territory Mapping of the Cerebral Arteries With Regional Perfusion MRI
Jeroen Hendrikse, Jeroen van der Grond, Hanzhang Lu et al. · Stroke · 2004 · 164 citations