Aerosol Science and Technology · 1998 · 15 citations · 8 references
EngineeringEnvironmental EngineeringAir PollutionAtmospheric ScienceInstrumentationAerodynamic TestingAerosol Particle EngineeringAerosol ScienceAir Pollution InstrumentationParticulate MatterWind-tunnel TestingAerodynamic Particle Size DistributionInlet AerodynamicsParticle Size Distribution MeasurementPm10Sampling
ABSTRACT A new PM10 inlet for a beta-gauge sampler was designed based on the particle cup impactor concept. The performance of the inlet was evaluated at near-zero wind velocity in the test chamber and at the wind velocities of 2 and 8 km/hr in the wind tunnel. The performance indicated that particles with aerodynamic diameters of 10 μm or larger were collected in the particle cup and the inlet proved to meet the basic requirement of PM10 sampling.
8