Erythrocyte and Reticulocyte Indices on the LH 750 as Potential Markers of Functional Iron Deficiency

16

Citations

22

References

2010

Year

Abstract

Reticulocyte hemoglobin content (CHr) and percentage of hypochromic cells (%Hypo) are restricted to the Siemens analysers. The aims of the study were to investigate the correlation of Red cells size factor (RSf) and low Hemoglobin density (LHD%), reported by Beckman-Coulter analysers, with CHr and %Hypo in the assessment of iron status in the presence of inflammation. 381 samples were run on both LH 750 (Beckman-Coulter) and Advia 2120 (Siemens) analysers. Correlation between parameters were calculated and the diagnostic performance of the new parameters was assessed. Results. Correlation between RSf and CHr, r = 0.85. ROC curve analysis for RSf in the diagnosis of iron restricted erythropoiesis defined as CHr < 28 pg: AUC 0.983; Cutoff 91.1%; Sensitivity 98.8%; Specificity 89.6% Correlation between LHD% and %Hypo, r = 0.869. ROC curve analysis for LHD% in the diagnosis of iron deficiency defined by %Hypo >5%: AUC 0.954; Cut off 6.0%; Sensitivity 96.6%; Specificity 83.2% Conclusions. RSf and LHD% could be reliable parameters for the study of iron metabolism status.