Speech Perception as a Function of Electrical Stimulation Rate: Using the Nucleus 24 Cochlear Implant System

305

Citations

37

References

2000

Year

Abstract

For the speech processing strategy, implant system, and subjects evaluated in this study, the group results indicated that the use of electrical stimulation rates higher than 250 pps/ch (up to 1615 pps/ch) generally provided no significant improvement to speech comprehension. However, individual results indicated that perceptual benefits could be obtained by adjusting rate of stimulation optimally to suit each subject. Results from one subject, together with tinnitus problems arising from high-rate stimulation for another subject, indicated that high rates of stimulation may in fact be undesirable for some subjects.