The application of phase shifts in NMR for flow measurement

David Firmin, G. L. Nayler, Philip J. Kilner, D B Longmore

Magnetic Resonance in Medicine · 1990 · 266 citations · 19 references

Concepts

TL;DR

Phase‑shift NMR imaging has evolved from direct phase mapping and Fourier flow imaging to magnitude‑based methods, and this review outlines the historical development and common error sources affecting these techniques. The study aims to discuss the principal errors that affect the accuracy of flow imaging techniques, focusing on phase‑mapping methods used at the authors’ institution. Experimental results demonstrate the magnitude of these errors and propose methods to remove or minimize them. © 1990 Academic Press, Inc.

Abstract

Abstract A brief overview of the history of the application of phase shifts in NMR, and in particular NMR imaging, is presented. The imaging methods include direct phase mapping, Fourier flow imaging (where the flow data are Fourier transformed into one dimension of an image), and alternative methods, where flow‐related phase shifts are utilized for flow measurement from the magnitude of the signal. A discussion then follows of the principal errors that can affect the accuracy of the various flow imaging techniques, with particular reference to the phase mapping methods that have been used extensively in our institution. The results from a number of experiments are included to illustrate the extent of the errors and methods of removing or minimizing these effects are Suggested. © 1990 Academic Press, Inc.