EngineeringNuclear EngineeringRadiation DetectionRadiation DosimetryRadiation MeasurementDetector SystemsDetector TechnologyGamma-ray DetectorsRadiation DetectorsRadiation Dose MeasurementRadiation InstrumentationDetector DesignDetector Efficiency CalibrationDetectorsDose Rate CalibrationField Instrument CalibrationField MeasurementsGamma-ray SpectrometryRadiation Detector Calibration
Calibration factors to convert a measured full-absorption peak count rate to activity in the soil and dose rate in air are given for Ge detectors that are used for field measurements of radionuclides. The appropriate factors for a given detector are derived using three primary parameters: the manufacturer's quoted efficiency at 1332 keV relative to a 7.6 cm (3 in.) long by 7.6 cm (3 in.) diameter NaI(Tl) detector, the detector's orientation in the field (up or down) and the Ge crystal length/diameter ratio. The accuracy of the results obtained by using this simplified calibration technique is estimated to be 10-15%.