EngineeringBioengineeringBiomedical EngineeringBioinstrumentationBiomedical Signal ProcessingNeural EngineeringBiomedical InstrumentationElectroencephalogram Signal ProcessingElectroencephalographyElectroencephalography In EpilepsyNeural Recording TechnologiesWireless ConnectivityElectroencephalogram AcquisitionElectroencephalography InstrumentationElectrophysiological RecordingsIntracranial EegPortabilityPower ConsumptionTransceiver DesignWireless Neural Recording
In this paper, we present system design of remote data recording for epileptic and stroke patients. We report wireless recording system combining near infra-red spectrometry (NIRS) in a comprehensive non-invasive evaluation and both electroencephalographic (EEG), and intracerebral EEG (icEEG) recording in presurgical evaluation. A Bluetooth and dual radio links were introduced for these recording. The Bluetooth-based device was embedded in a non-invasive multichannel EEG-NIRS system for easy portability and long term monitoring. On the other hand, the recorded icEEG signals from up to 128 channels through intracerebral electrodes are transmitted wirelessly to a remote base station. This transmitter front-end contains preamplifier, proper filtering, data converter and power amplifier. The RF back-end is based on commercial transceiver operated in the Medical Implant Communication Service (MICS) 400 MHz band. Power consumption of the front-end and transmitter are 0.75mW and 15mW, respectively. The proposed remote monitoring systems are validated in vitro using data recording from human patients.
3
Étienne Lareau, Frédéric Lesage, Philippe Pouliot et al. · Journal of Biomedical Optics · 2011 · 66 citations · Full text