MedicineNuclear EngineeringRadiologyDiagnostic ImagingDiagnostic RadiologyRadiation DosimetryImage QualityMedical ImagingNuclear Medicine ImagingProton TherapyRadiation Imaging SystemsRadiographic ImagingDensityDepth Dose DistributionHigh-resolution ImagingImaging ResolutionProton Dose DistributionRadiographic Image QualitySpatial Resolution AnalysisSpatial Resolution Measurement
The image quality of proton radiography using the method of beam energy modulation was studied to look into its practical uses. Two different depth-dose distributions generated by modulation were applied to investigate their effects on the density and spatial resolutions of the radiographic image. A steeper slope was found to provide higher resolution for the matching thickness of a phantom. The image was taken on a scintillation screen, and the distance between the phantom and the screen was a sensitive parameter on the resolution. For a beam with a range of 1.2 cm in Lucite, high-resolution images were attainable in the whole range. The resolution of proton images for different kinds of phantoms was examined in comparison with x-ray images as well as images simulated by a Monte Carlo code MCNPX. For a longer range of 18 cm, images attained by simulations indicated that density resolution is better maintained compared to spatial resolution, which is deteriorated by multiple Coulomb scattering.
8
Radiological Use of Fast Protons
Robert R. Wilson · Radiology · 1946 · 1.4K citations
Natural Sciences, Physics, Radiology +14
Proton radiography as a tool for quality control in proton therapy
Uwe Schneider, Eros Pedroni · Medical Physics · 1995 · 199 citations
Andreas Koehler · Science · 1968 · 140 citations
First proton radiography of an animal patient
Uwe Schneider, Jürgen Besserer, Peter Pemler et al. · Medical Physics · 2004 · 112 citations