Magnetic Resonance in Medicine · 2000 · 586 citations · 19 references
MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Pulse SequencesMagnetic Resonance Signal FormationMedical ImagingMedical Imaging InstrumentationB0 Field Inhomogeneity CorrectionEcho Planar ImagingEddy Current AnalysisEddy Current CorrectionGradient Echo SequencesMagnetic Field MappingStatic Magnetic Field
Improvements in Bo mapping and shimming were achieved by measuring the static field information in multiple subsequent echoes generated by an asymmetric echo-planar readout gradient train. With careful compensation, eddy current effects were shown to affect the adjustment of the shim coils minimally. In addition to reducing the time required for field mapping by two-fold, the sensitivity was simultaneously optimized irrespective of the prevalent T2* present, thereby minimizing the error of the static field measurement to below 0.1 Hz. With adiabatic low flip-angle excitation, the time required for field mapping was below 1 second.