Clinical Physics and Physiological Measurement · 1992 · 39 citations · 13 references
MedicineOrthopaedic SurgeryRadiologyBone MetabolismUltrasonographyUltrasoundBone HealthDiagnostic UltrasonographyMusculoskeletal UltrasonographyMusculoskeletal UltrasoundUltrasonicsUltrasound ImagingBone Density EstimationBone Mineral DensityCorrelationMineral Content EstimationMusculoskeletal Ultrasound TechniqueRadiodensity MeasurementUltrasonic AttenuationUltrasonic Imaging
The reproducibility of two methods of measuring broadband ultrasonic attenuation (BUA) in the calcaneus have been studied. An improvement in reproducibility in vivo from 9.6% to 2.8% between old and new techniques has been observed. Measurements of the calcaneus using BUA were correlated with measurements of bone mineral density measured by dual energy x-ray absorptiometry in the lumbar spine, femur and total body and bone mineral content in the distal and proximal forearm measured by single photon absorptiometry. For the older BUA technique the correlation coefficients ranged between r = 0.27 and r = 0.34. For the newer BUA technique the correlation coefficients ranged between r = 0.49 and r = 0.62 and were all significant (P less than 0.001).
13
Measurement of Bone Mineral in vivo: An Improved Method
John Cameron, James A. Sorenson · Science · 1963 · 832 citations · Full text
Medicine, Radiology, Bone Metabolism +14
Quantification of Bone Mineralization Using Computed Tomography
P. Rüegsegger, U. Elsasser, M. Anliker et al. · Radiology · 1976 · 197 citations
Medicine, Oral Radiology, Radiology +17
Phillip J. Rossman, Cristian Meșină, James A. Sorenson et al. · Clinical Physics and Physiological Measurement · 1989 · 149 citations