Automation of an ultrasound velocity measurement system in high-pressure liquids

Zhong Ding, J Alliez, Christian Boned, P. Xans

Measurement Science and Technology · 1997 · 22 citations · 4 references

Concepts

TL;DR

The study automated an ultrasound velocity measurement system for liquids to handle temperature and pressure variations. The automated system monitors temperature and pressure, generates and receives signals, acquires measurement points, identifies signals using a reference, measures ultrasound travel times, stores and processes results, and includes a calibration procedure. The system saves time relative to manual methods and achieves 0.1 % average and 0.25 % maximum deviation from literature values for octane and pentane.

Abstract

A system for the measurement of ultrasound velocity in liquids was automated to cope with variations in temperature and pressure . The following problems had to be solved: supervision and monitoring of the achievement of the required pressure and temperature, creation and reception of the signal, acquisition of the measurement points, identification of the signal received on the basis of the characteristics of a reference signal, measurement of ultrasound travel times, memorization and processing of the results. The system enables substantial savings of time compared with manual methods. The system calibration procedure is also described. Measurements of octane and pentane (which had been studied previously) are used to verify the validity of the system developed and to determine its degree of accuracy (absolute average deviation from literature data 0.1% and maximum deviation from literature data 0.25%).

References

4