Computed diffusion-weighted MRI for prostate cancer detection: the influence of the combinations of<i>b</i>-values

Yoshiko Ueno, Satoru Takahashi, Yoshiharu Ohno, K. Kitajima, Masao Yui, Yoshimori Kassai, Fumi Kawakami, Hideaki Miyake, Kazuro Sugimura

British Journal of Radiology · 2015 · 39 citations · 13 references

DOIFull text

Open access

Concepts

Abstract

OBJECTIVE: To evaluate the influence of the combinations of b-values on computed diffusion-weighted images (cDWIs) for prostate cancer (PCa) detection at b = 2000 s mm(-2). METHODS: Diffusion-weighted imaging (DWIs) for 31 patients with PCa (65.2 ± 7.1 years) were obtained pre-operatively at different b-values (0, 100, 500, 1000 and 2000 s mm(-2)) on a 3-T MRI. cDWIs at b = 2000 were generated by using six b-value combinations: 0-100 s mm(-2) (cDWI0-100); 0-500 s mm(-2) (cDWI0-500); 100-500 s mm(-2) (cDWI100-500); 0-1000 s mm(-2) (cDWI0-1000); 100-1000 s mm(-2) (cDWI100-1000); and 500-1000 s mm(-2) (cDWI500-1000). These cDWIs and measured DWIs with b = 2000 s mm(-2) (mDWI2000) were evaluated in this setting. To assess image quality for each DWI, contrast ratios (CRs) of cancerous and non-cancerous lesions were evaluated. To compare the detectability of PCa for each DWI, receiver operating characteristic analysis was used. RESULTS: CRs of all cDWIs were significantly higher than those of mDWI2000 (p < 0.05). Areas under the curve of cDWI0-100 (0.62) and cDWI0-500 (0.65) were significantly smaller (p < 0.05) than those of others (cDWI100-500, 0.72; cDWI0-1000, 0.73; cDWI100-1000, 0.71; cDWI500-1000, 0.74; mDWI2000, 0.72). CONCLUSION: The combinations of b-values influenced image quality and diagnostic ability of cDWIs for PCa detection. The combinations of b ≥ 100 and b ≥ 500 s mm(-2), as well as b = 0 and b = 1000 s mm(-2), were optimal in this study. ADVANCES IN KNOWLEDGE: For generating the useful cDWI for PCa detection, radiologists should take care of the combination of b-values when including low b-values.

References

13