Magnetic Resonance in Medicine · 1997 · 154 citations · 5 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Pulse SequencesMedical ImagingNoiseNoise PredictionAcoustic NoiseAcoustic Noise ModelingFast Spin EchoGradient Echo SequencesGradient NoiseSpin EchoSpin Echo SequencesWhite Noise
Gradient acoustic noise has been measured and characterized for an epoxy-potted, shielded gradient assembly in a 1.5 T MRI system. Noise levels vary by 10 dB or more as a function of longitudinal position in the scanner and reflect the pattern of forces applied to the gradient assembly. The noise level increases slightly (1-3 dB) with a patient in the scanner. The spectrum of the noise is similar (but not identical) to the spectrum of the input signal. A gradient-pulse-to-acoustic-noise transfer function was obtained by using a white noise voltage input to the gradient system. The transfer function enabled us to accurately predict acoustic noise output for a pulse sequence consisting of a series of trapezoidal pulses on a single axis and for a clinical fast spin echo sequence with gradients present on all three axes.