Optics Express · 2013 · 30 citations · 28 references
EngineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsDiagnostic ImagingContrast Enhancement StrategiesImagingOptical ImagingAccuracyContrast-to-noise Ratio EstimationLaser Speckle ImagingLight IntensityLocal Contrast EnhancementPixel Size EffectsSpeckle ContrastStatic Contrast
In laser speckle contrast imaging, it was usually suggested that speckle size should exceed two camera pixels to eliminate the spatial averaging effect. In this work, we show the benefit of enhancing signal to noise ratio by correcting the speckle contrast at small speckle size. Through simulations and experiments, we demonstrated that local speckle contrast, even at speckle size much smaller than one pixel size, can be corrected through dividing the original speckle contrast by the static speckle contrast. Moreover, we show a 50% higher signal to noise ratio of the speckle contrast image at speckle size below 0.5 pixel size than that at speckle size of two pixels. These results indicate the possibility of selecting a relatively large aperture to simultaneously ensure sufficient light intensity and high accuracy and signal to noise ratio, making the laser speckle contrast imaging more flexible.
28
Dynamic Imaging of Cerebral Blood Flow Using Laser Speckle
Andrew K. Dunn, Hayrünnisa Bolay, Michael A. Moskowitz et al. · Journal of Cerebral Blood Flow & Metabolism · 2001 · 824 citations · Full text
Medicine, Neurology, Neuroscience +17
Robust flow measurement with multi-exposure speckle imaging
Ashwin B. Parthasarathy, W. James Tom, Ashwini Gopal et al. · Optics Express · 2008 · 365 citations · Full text