Publication | Closed Access
Optical eye tracking system for noninvasive and automatic monitoring of eye position and movements in radiotherapy treatments of ocular tumors
16
Citations
20
References
2012
Year
A noninvasive eye tracking system based on infrared 3-D video-oculographic techniques is proposed for the automatic monitoring of eye position and orientation in external beam radiotherapy of ocular tumors. The presented method can be applied for the real-time estimation of lesion position and tumor-beam misalignments, allowing automatic patient setup and eye movement gated treatments. A prototypal eye tracker was developed and tested on five subjects, achieving gaze estimation errors of 0.5° and eye monitoring frequencies of 125 Hz. The proposed application can potentially improve quality and efficacy of ocular radiotherapy treatments, currently based on invasive, qualitative, and manual control procedures.